Alternative Biomedical Solutions
sells a complete line of
Quality
Refurbished Olympus Chemistry Immuno
Systems.
When you
purchase from ABS you get:
-
Quality Refurbished Systems at
Significant
Savings
-
Installation and onsite Operator
Training
-
Warranty – 12 months parts, labor &
travel
- Technical Support
-
Onsite Response – 24 to 48 hour
-
Options – 24/7 and Business Hours
|
|
Analytical system
|
Fully automated, for
routine, STAT, urine, Fluid samples |
|
Analytical methods
|
Clinical chemistry and
immunological parameters. End point,
kinetic assays, fixed-time-kinetics,
optionally ISE |
|
Simultaneously processed
analytes
|
48 photometric tests, 51
with ISE |
|
Throughput
|
800 photometric
tests/hour; maximum of 1200 with ISE
|
|
Sample feeder
|
Racks with 10 samples
each (bar codes on primary tubes and on
racks); capacity of 150 samples;
continuous loading |
|
Sample tubes
|
In primary and secondary
tubes; diameter between 11.5 and 16.5
mm; height between 55 and 102 mm
|
|
STAT samples
|
Up to 18 positions for
STAT sampleswith bar codes |
|
Sample volume
|
2 – 50 µl in 1 µl steps
(1 – 50 µl for repeats)
|
|
Reagent supply
|
refrigerated (4 °C - 12
°C); 48 positions for R1/R2 |
|
Reagent volume
|
1. reagent: 25 – 300 µl;
2. reagent: 25 – 300 µl; (in 1 µl steps)
|
|
Total reaction volume
|
152 – 550 µl
|
|
Reaction cuvette
|
Quartz cuvettes
|
|
Reaction time
|
Up to 8 minutes, 15
seconds |
|
Reaction temperature
|
37 °C |
|
Mixing method
|
With rotating paddles
after dispensing sample and reagent
|
|
Photometry system
|
Direct assay through the
reaction cuvette (0 – 2.5 OD) mono,
bichromatic measurements possible
|
|
Wavelength
|
13 different wave lengths
between 340 – 800 nm |
|
Cuvette cleaning
|
Comprehensive cleaning
with detergents |
|
Calibration
|
linear and non linear
(linear interpolation) |
|
Data processing
|
calculated parameters,
blank value correction and correlation.
Definition of normal
value range, measuring range,
pathological values and repeats. Quality
controls,
printout of abnormal values, pending
sample work list and repeat run sample
list |
|
Test requisition
|
Individual and profile
test requisition via on line, mouse
or functions keys |
|
Water supply /waste |
automated supply with
deionised water (40 l / hour); waste up
to 100 mm height |
|
Dimensions (W xHxD) (mm)
|
ANL: 1680 x 1210 x 800;
DPR: 700 x 1430 x 700 |
|
Weight (kg) |
ANL: 480; PC: 70 |
|
On-line |
full uni- and
bidirectional communication possible |
|
Power supply |
240 V ±10%; 4 kVA |