Analytical System |
Discrete, twin reaction lines |
|
Analytical method |
End point assay, rate assay,
fixed assay. |
|
Simultaneous analysis items
|
16, 24, 32 |
|
Max processing capacity
|
7,920 chemistry results/hour (24
chemistries x 330 samples/hour). |
|
Sample feeder |
Rack system with a capacity of
10 samples per rack, with
bar-coded sample readout
capability. Accommodates 300
samples. |
|
Sample vessel |
14mm x 55mm(L) glass or plastic
13—16.5mmx55—102mm(L) primary
tubes may also be used. |
|
Emergency Samples |
Interruptive analysis possible
by loading emergency sample in
the red rack. |
|
Repeat run samples
|
Repeat run is possible by
loading repeat run samples in
the orange rack. |
|
Sample volume |
3—15ul in 1ul increments (1—15ul
for repeat run). 4-test split
dispensing. |
|
Reagent storage |
Refrigerated |
|
Reagent Volume |
1st reagent: 50—450ul
in 1ul
increments
2nd reagent: 50—350ul
in 1ul
increments
2nd reagent is
dispensed 3.6 minutes after
sample is dispensed |
|
Applicable reagents
|
Commercially available reagents |
|
Reaction line |
Twin reaction lines (turntable).
Dry bath incubation method.
|
|
Reaction time
|
8
minutes and 31 seconds |
|
Reaction temperature
|
37°C |
|
Mixing method
|
Rotating action from the
stirring bars after dispensing
the sample and after dispensing
the second reagent. |
|
Photometry System
|
Direct
photometry in reaction cuvettes.
Required reaction mixture
volume: 250ul, light path: 6mm.
One or two-wavelength photometry
with the diffraction grating.
|
|
Cuvette washing |
Washing and drying within the
glass cuvette wheel |
|
Calibration curve selection
|
Linear and sectional lines
Automatic: 5 points max.;
Manual: 7 points max. |
|
Quality Control |
By
loading quality control samples
in the green rack (3 types for
each test, total of 8 samples
types). Daily and day-to-day
mean (X), range ®, standard
deviation (SD) and coefficient
of variation (CV) and
demographics. Alarm for
exceeding control limit. |
|
Data processing |
Test
ratio calculations, sample blank
correction, and correlatire
correction; normal, panic,
dynamic and repeat run range
judgment; reagent absorbance and
reaction mixture absorbance
check; abnormal value, pending
list and repeat run wertesheet
output. |
|
Test item selection
|
Single or/and panel
99 different panel
definable
Requisition of 4,000 routine and
200 emergency samples possible.
|
|
Input/output |
Keyboard entry and on-line
requisition. CRT, high-speed and
online |
|
Water supply/drainage
|
Automatic supply of deionized
water (30L/unit hr.)Drainage
outlet not higher than 100mm
above the floor level
|
|
Ambient Conditions
|
temperature: 15--30°C,
within +- 2°C fluctuation during
measurement, Relative humidity:
40—80% (non-condensing
atmosphere). |